:Weston cell

{{Short description|Wet chemical cell that produces a precise voltage}}

File:Westoncell Volt.png

File:Weston cell - Hanford B Reactor - Benton County, Washington.jpg

File:Standart cell.jpg

The Weston cell or Weston standard cell is a wet-chemical cell that produces a highly stable voltage suitable as a laboratory standard for calibration of voltmeters. Invented by Edward Weston in 1893, it was adopted as the International Standard for EMF from 1911 until superseded by the Josephson voltage standard in 1990.

Chemistry

The anode is an amalgam of cadmium with mercury with a cathode of pure mercury over which a paste of mercurous sulfate and mercury is placed. The electrolyte is a saturated solution of cadmium sulfate, and the depolarizer is a paste of mercurous sulfate.

As shown in the illustration, the cell is set up in an H-shaped glass vessel with the cadmium amalgam in one leg and the pure mercury in the other. Electrical connections to the cadmium amalgam and the mercury are made by platinum wires fused through the lower ends of the legs.

;Anode reaction: Cd(s) → Cd2+(aq) + 2e

;Cathode reaction: (Hg+)2{{chem|SO|4|2−}}(s) + 2e → 2Hg(l) + {{chem|SO|4|2−}}(aq)

Reference cells must be applied in such a way that no current is drawn from them.

Characteristics

The original design was a saturated cadmium cell producing a {{val|1.018638|ul=V}} reference and had the advantage of having a lower temperature coefficient than the previously used Clark cell.{{Cite book |last=Northrop |first=Robert |url=https://archive.org/details/introduction_to_instrumentation_and_measurements |title=introduction to instrumentation and measurements |date=2005 |isbn=978-1-4200-5785-0 |edition=2 |pages=14|publisher=CRC Press }}

One of the great advantages of the Weston normal cell is its small change of electromotive force with change of temperature. At any temperature {{mvar|t}} between {{val|0|u=°C}} and {{val|40|u=°C}},

:{{math|Et/V {{=}} E20/V − {{val|0.0000406}} (t/°C − 20) − {{val|0.00000095}} (t/°C − 20)2 + {{val|0.00000001}} (t/°C − 20)3}}.

This temperature formula was adopted by the London conference of 1908.{{cite journal

| title = Electric units and standards

| journal = Circular of the National Bureau of Standards

| volume = 1916

| issue = 58

| pages = 39

| publisher = USA Government Printing Office

| location = Washington, D.C.

| date = 25 September 1916

| language = english

| url = https://books.google.com/books?id=_EsbAAAAMAAJ&pg=RA2-PA39

| access-date = 12 July 2016}}

The temperature coefficient can be reduced by shifting to an unsaturated design, the predominant type today. However, an unsaturated cell's output decreases by some 80 microvolts per year, which is compensated by periodic calibration against a saturated cell.

See also

References

{{Reflist}}

Literature

  • Practical Electricity by W. E. Ayrton and T. Mather, published by Cassell and Company, London, 1911, pp 198–203
  • {{US patent|494827}}, "Voltaic cell"
  • [https://www.nist.gov/calibrations/upload/mn84.pdf Standard Cells, Their Construction, Maintenance, and Characteristics] by Walter J. Hamer, National Bureau of Standards Monograph 84, January 15, 1965.