Electron energy loss spectroscopy
{{short description|Form of microscopy using an electron beam}}
{{redirect|EELS|other uses|Eel (disambiguation)}}
File:Electron energy loss spectrum feature overview.svg
Electron energy loss spectroscopy (EELS) is a form of electron microscopy in which a material is exposed to a beam of electrons with a known, narrow range of kinetic energies. Some of the electrons will undergo inelastic scattering, which means that they lose energy and have their paths slightly and randomly deflected. The amount of energy loss can be measured via an electron spectrometer and interpreted in terms of what caused the energy loss. Inelastic interactions include phonon excitations, inter- and intra-band transitions, plasmon excitations, inner shell ionizations, and Cherenkov radiation. The inner-shell ionizations are particularly useful for detecting the elemental components of a material. For example, one might find that a larger-than-expected number of electrons comes through the material with 285 eV less energy than they had when they entered the material. This is approximately the amount of energy needed to remove an inner-shell electron from a carbon atom, which can be taken as evidence that there is a significant amount of carbon present in the sample. With some care, and looking at a wide range of energy losses, one can determine the types of atoms, and the numbers of atoms of each type, being struck by the beam. The scattering angle (that is, the amount that the electron's path is deflected) can also be measured, giving information about the dispersion relation of whatever material excitation caused the inelastic scattering.{{cite journal |last1=Egerton |first1=R. F. |year=2009 |title=Electron energy-loss spectroscopy in the TEM |journal=Reports on Progress in Physics |volume=72 |issue=1 |page=016502 |bibcode=2009RPPh...72a6502E |doi=10.1088/0034-4885/72/1/016502|s2cid=120421818 }}
History
The technique was developed by James Hillier and RF Baker in the mid-1940s{{cite journal |last2=Hillier |first1=J. |last1=Baker |first2=R. F. |date=September 1944 |title=Microanalysis by means of electrons |journal=J. Appl. Phys. |volume=15 |issue=9 |pages=663–675 |doi=10.1063/1.1707491 |bibcode=1944JAP....15..663H}} but was not widely used over the next 50 years, only becoming more widespread in research in the 1990s due to advances in microscope instrumentation and vacuum technology. With modern instrumentation becoming widely available in laboratories worldwide, the technical and scientific developments from the mid-1990s have been rapid. The technique is able to take advantage of modern aberration-corrected probe forming systems to attain spatial resolutions down to ~0.1 nm, while with a monochromated electron source and/or careful deconvolution the energy resolution can reach units of meV.{{Cite journal |last=Plotkin-Swing |first=B |last2=Mittelberger |first2=A |last3=Haas |first3=B |last4=Idrobo |first4=J C |last5=Graner |first5=B |last6=Dellby |first6=N |last7=Hotz |first7=M T |last8=Meyer |first8=C E |last9=Quillin |first9=S C |last10=Krivanek |first10=O L |last11=Lovejoy |first11=T C |date=2023-07-22 |title=Ultra-high Energy Resolution EELS and 4D STEM at Cryogenic Temperatures |url=https://academic.oup.com/mam/article/29/Supplement_1/1698/7229073 |journal=Microscopy and Microanalysis |language=en |volume=29 |issue=Supplement_1 |pages=1698–1699 |doi=10.1093/micmic/ozad067.875 |issn=1431-9276}} This has enabled detailed measurements of the atomic and electronic properties of single columns of atoms, and in a few cases, of single atoms.{{cite journal|last1=Ramasse|first1=Quentin M.|last2=Seabourne|first2=Che R.|last3=Kepaptsoglou|first3=Despoina-Maria|last4=Zan|first4=Recep|last5=Bangert|first5=Ursel|author-link5=Ursel Bangert|last6=Scott|first6=Andrew J.|date=October 2013|title=Probing the Bonding and Electronic Structure of Single Atom Dopants in Graphene with Electron Energy Loss Spectroscopy|journal=Nano Letters|volume=13|issue=10|pages=4989–4995|bibcode=2013NanoL..13.4989R|doi=10.1021/nl304187e|issn=1530-6984|pmid=23259533|s2cid=68082 }}{{cite journal |last1=Tan |first1=H. |last2=Turner |first2=S. |last3=Yücelen |first3=E. |last4=Verbeeck |first4=J. |last5=Van Tendeloo |first5=G. |date=September 2011 |title=2D atomic mapping of oxidation states in transition metal oxides by scanning transmission electron microscopy and electron energy-loss spectroscopy |journal=Phys. Rev. Lett. |volume=107 |issue=10 |page=107602 |doi=10.1103/PhysRevLett.107.107602|pmid=21981530 |bibcode = 2011PhRvL.107j7602T |hdl=10067/912650151162165141 |url=http://repository.tudelft.nl/islandora/object/uuid%3Ab99dc10f-5f80-44b3-b258-9f6326c33d90/datastream/OBJ/view |hdl-access=free }}
Comparison with EDX
EELS is spoken of as being complementary to energy-dispersive x-ray spectroscopy (variously called EDX, EDS, XEDS, etc.), which is another common spectroscopy technique available on many electron microscopes. EDX excels at identifying the atomic composition of a material, is quite easy to use, and is particularly sensitive to heavier elements. EELS has historically been a more difficult technique but is in principle capable of measuring atomic composition, chemical bonding, valence and conduction band electronic properties, surface properties, and element-specific pair distance distribution functions.{{sfn|Egerton|1996}} EELS tends to work best at relatively low atomic numbers, where the excitation edges tend to be sharp, well-defined, and at experimentally accessible energy losses (the signal being very weak beyond about 3 keV energy loss). EELS is perhaps best developed for the elements ranging from carbon through the 3d transition metals (from scandium to zinc).Ahn C C (ed.) (2004) Transmission electron energy loss spectrometry in material science and the EELS Atlas, Wiley, Weinheim, Germany, {{doi|10.1002/3527605495}}, {{ISBN|3527405658}} For carbon, an experienced spectroscopist can tell at a glance the differences between diamond, graphite, amorphous carbon, and "mineral" carbon (such as the carbon appearing in carbonates). The spectra of 3d transition metals can be analyzed to identify the oxidation states of the atoms.{{Cite journal |doi=10.1016/j.micron.2006.06.017 |pmid=16962785 |volume=38 |issue=3 |pages=224–230 |last=Riedl |first=T. |author2=T. Gemming |author3=W. Gruner |author4=J. Acker |author5=K. Wetzig |title=Determination of manganese valency in La1−xSrxMnO3 using ELNES in the (S)TEM |journal=Micron |date=April 2007}} Cu(I), for instance, has a different so-called "white-line" intensity ratio than Cu(II) does. This ability to "fingerprint" different forms of the same element is a strong advantage of EELS over EDX. The difference is mainly due to the difference in energy resolution between the two techniques (~1 eV or better for EELS, perhaps a few tens of eV for EDX).
Variants
File:Electron energy loss spectroscopy coreloss lsmo.svg.]]
There are several basic flavors of EELS, primarily classified by the geometry and by the kinetic energy of the incident electrons (typically measured in kiloelectron-volts, or keV). Probably the most common today is transmission EELS, in which the kinetic energies are typically 100 to 300 keV and the incident electrons pass entirely through the material sample. Usually this occurs in a transmission electron microscope (TEM), although some dedicated systems exist which enable extreme resolution in terms of energy and momentum transfer at the expense of spatial resolution.{{cn|date=February 2024}}
Other flavors include reflection EELS (including reflection high-energy electron energy-loss spectroscopy (RHEELS)), typically at 10 to 30 keV, and aloof EELS (sometimes called near-field EELS), in which the electron beam does not in fact strike the sample but instead interacts with it via the long-ranged Coulomb interaction. Aloof EELS is particularly sensitive to surface properties but is limited to very small energy losses such as those associated with surface plasmons or direct interband transitions.{{cn|date=February 2024}}
Within transmission EELS, the technique is further subdivided into valence EELS (which measures plasmons and interband transitions) and inner-shell ionization EELS (which provides much the same information as x-ray absorption spectroscopy, but from much smaller volumes of material). The dividing line between the two, while somewhat ill-defined, is in the vicinity of 50 eV energy loss.
Instrumental developments have opened up the ultra-low energy loss part of the EELS spectrum, enabling vibrational spectroscopy in the TEM.{{Cite journal|last1=Krivanek|first1=Ondrej L.|last2=Lovejoy|first2=Tracy C.|last3=Dellby|first3=Niklas|last4=Aoki|first4=Toshihiro|last5=Carpenter|first5=R. W.|last6=Rez|first6=Peter|last7=Soignard|first7=Emmanuel|last8=Zhu|first8=Jiangtao|last9=Batson|first9=Philip E.|last10=Lagos|first10=Maureen J.|last11=Egerton|first11=Ray F.|date=2014|title=Vibrational spectroscopy in the electron microscope|journal=Nature|language=en|volume=514|issue=7521|pages=209–212|doi=10.1038/nature13870|pmid=25297434|issn=0028-0836|bibcode=2014Natur.514..209K|s2cid=4467249 }} Both IR-active and non-IR-active vibrational modes are present in EELS.{{Cite journal|last1=Venkatraman|first1=Kartik|last2=Levin|first2=Barnaby D.A.|last3=March|first3=Katia|last4=Rez|first4=Peter|last5=Crozier|first5=Peter A.|date=2019|title=Vibrational spectroscopy at atomic resolution with electron impact scattering|url=https://www.nature.com/articles/s41567-019-0675-5|journal=Nature Physics|language=en|doi=10.1038/s41567-019-0675-5|arxiv=1812.08895|volume=15|issue=12|pages=1237–1241|s2cid=119452520 }}
EEL spectrum
The electron energy loss (EEL) spectrum (sometimes spelled EELS spectrum) can be roughly split into two different regions: the low-loss spectrum (up until about 50 eV in energy loss) and the high-loss spectrum. The low-loss spectrum contains the zero-loss peak (signal from all the electrons which did not loose a measurable energy) as well as the phonon{{Cite journal |last=Krivanek |first=O. L. |last2=Dellby |first2=N. |last3=Hachtel |first3=J. A. |last4=Idrobo |first4=J. -C. |last5=Hotz |first5=M. T. |last6=Plotkin-Swing |first6=B. |last7=Bacon |first7=N. J. |last8=Bleloch |first8=A. L. |last9=Corbin |first9=G. J. |last10=Hoffman |first10=M. V. |last11=Meyer |first11=C. E. |last12=Lovejoy |first12=T. C. |date=2019-08-01 |title=Progress in ultrahigh energy resolution EELS |url=https://linkinghub.elsevier.com/retrieve/pii/S0304399118303668 |journal=Ultramicroscopy |series=75th Birthday of Christian Colliex, 85th Birthday of Archie Howie, and 75th Birthday of Hannes Lichte / PICO 2019 - Fifth Conference on Frontiers of Aberration Corrected Electron Microscopy |volume=203 |pages=60–67 |doi=10.1016/j.ultramic.2018.12.006 |issn=0304-3991}} and plasmon peaks, and contains information about the band structure and dielectric properties of the sample. It is also possible to resolve the energy spectrum in momentum to directly measure the band structure. The high-loss spectrum contains the ionisation edges that arise due to inner shell ionisations in the sample. These are characteristic to the species present in the sample, and as such can be used to obtain accurate information about the chemistry of a sample.{{Cite journal|last=Hofer|first=F. |display-authors=etal |date=2016|title=Fundamentals of electron energy loss spectroscopy|journal=IOP Conference Series: Materials Science and Engineering|volume=109|pages=012007|doi=10.1088/1757-899X/109/1/012007|doi-access=free}}
Typically, EEL spectra are susceptible to noise, especially for measurements of beam sensitive materials, such as polymers or biological specimen, requiring limited acquisition times. The two major noise contributions are Poisson noise arising from the quantized nature of the beam electrons and Gaussian distributed detector noise.{{Cite journal | last1 = Zietlow | first1 = C. | last2 = Lindner | first2 = J.K.N. | doi = 10.1038/s41598-025-85982-4 | title = An applied noise model for scintillation-based CCD detectors in transmission electron microscopy | journal = Scientific Reports | volume = 15 | pages = 3815 | year = 2025 | pmid = 39885260 | url= https://www.nature.com/articles/s41598-025-85982-4| pmc = 11782531 }} As EEL spectra are usually measured on CCD or direct electron detectors, where multiple pixels of a pixel-column are summed to create a spectrum out of a 2D pixel array, the noise statistics of such spectra is altered compared to regular 2D images. Due to the image formation process, especially on scintillation-based CCD detectors, the Poisson noise is also heavily correlated by the detector.{{Cite journal | last1 = Zietlow | first1 = C. | last2 = Lindner | first2 = J.K.N. | doi = 10.1016/j.ultramic.2024.114101 | title = An applied noise model for low-loss EELS maps | journal = Ultramicroscopy | volume = 270 | pages = 114101 | year = 2025 | pmid = 39823700 | url= https://www.sciencedirect.com/science/article/pii/S0304399124001803| doi-access = free }}
Thickness measurements
EELS allows quick and reliable measurement of local thickness in transmission electron microscopy.{{sfn|Egerton|1996}} The most efficient procedure is the following:{{cite journal |last1=Iakoubovskii |first1=K. |last2=Mitsuishi |first2=K. |last3=Nakayama |first3=Y. |last4=Furuya |first4=K. |year=2008 |title=Thickness measurements with electron energy loss spectroscopy |journal=Microscopy Research and Technique |volume=71 |issue=8 |pages=626–31 |doi=10.1002/jemt.20597 |pmid=18454473 |url=http://pubman.nims.go.jp/pubman/item/escidoc:1587374:1/component/escidoc:1587373/mrt626.pdf |citeseerx=10.1.1.471.3663 |s2cid=24604858 |access-date=2013-03-04 |archive-date=2017-09-22 |archive-url=https://web.archive.org/web/20170922235558/http://pubman.nims.go.jp/pubman/item/escidoc:1587374:1/component/escidoc:1587373/mrt626.pdf |url-status=dead }}
- Measure the energy loss spectrum in the energy range about −5..200 eV (wider better). Such measurement is quick (milliseconds) and thus can be applied to materials normally unstable under electron beams.
- Analyse the spectrum: (i) extract zero-loss peak (ZLP) using standard routines; (ii) calculate integrals under the ZLP (I0) and under the whole spectrum (I).
- The thickness t is calculated as mfp*ln(I/I0). Here mfp is the mean free path of electron inelastic scattering, which has been tabulated for most elemental solids and oxides.{{cite journal |last1=Iakoubovskii |first1=Konstantin |last2=Mitsuishi |first2=Kazutaka |last3=Nakayama |first3=Yoshiko |last4=Furuya |first4=Kazuo |year=2008 |title=Mean free path of inelastic electron scattering in elemental solids and oxides using transmission electron microscopy: Atomic number dependent oscillatory behavior |journal=Physical Review B |volume=77 |issue=10 |pages=104102 |bibcode=2008PhRvB..77j4102I |doi=10.1103/PhysRevB.77.104102 |url=http://pubman.nims.go.jp/pubman/item/escidoc:1587372:1/component/escidoc:1587371/Prb104102.pdf |access-date=2013-03-04 |archive-date=2016-03-03 |archive-url=https://web.archive.org/web/20160303215455/http://pubman.nims.go.jp/pubman/item/escidoc:1587372:1/component/escidoc:1587371/Prb104102.pdf |url-status=dead }}
The spatial resolution of this procedure is limited by the plasmon localization and is about 1 nm,{{sfn|Egerton|1996}} meaning that spatial thickness maps can be measured in scanning transmission electron microscopy with ~1 nm resolution.
Pressure measurements
The intensity and position of low-energy EELS peaks are affected by pressure. This fact allows mapping local pressure with ~1 nm spatial resolution.
- Peak shift method is reliable and straightforward. The peak position is calibrated by independent (usually optical) measurement using a diamond anvil cell. However, the spectral resolution of most EEL spectrometers (0.3-2 eV, typically 1 eV) is often too crude for the small pressure-induced shifts. Therefore, the sensitivity and accuracy of this method is relatively poor. Nevertheless, pressures as small as 0.2 GPa inside helium bubbles in aluminum have been measured.{{cite journal |last1=Taverna |first1=D. |last2=Kociak |first2=M. |last3=Stéphan |first3=O. |last4=Fabre |first4=A. |last5=Finot |first5=E. |last6=Décamps |first6=B. |last7=Colliex |first7=C. |year=2008 |title=Probing Physical Properties of Confined Fluids within Individual Nanobubbles |journal=Physical Review Letters |volume=100 |issue=3 |arxiv=0704.2306 |bibcode = 2008PhRvL.100c5301T |doi=10.1103/PhysRevLett.100.035301 |pmid=18232994 |page=035301|s2cid=4028240 }}
- Peak intensity method relies on pressure-induced change in the intensity of dipole-forbidden transitions. Because this intensity is zero for zero pressure the method is relatively sensitive and accurate. However, it requires existence of allowed and forbidden transitions of similar energies and thus is only applicable to specific systems, e.g., Xe bubbles in aluminum.{{cite journal |last1=Iakoubovskii |first1=Konstantin |last2=Mitsuishi |first2=Kazutaka |last3=Furuya |first3=Kazuo |year=2008 |title=Structure and pressure inside Xe nanoparticles embedded in Al |journal=Physical Review B |volume=78 |issue=6 |pages=064105 |bibcode=2008PhRvB..78f4105I |doi=10.1103/PhysRevB.78.064105 |url=http://pubman.nims.go.jp/pubman/item/escidoc:1587370:1/component/escidoc:1587369/Prb64105.pdf |access-date=2013-03-04 |archive-date=2020-07-31 |archive-url=https://web.archive.org/web/20200731004630/http://pubman.nims.go.jp/pubman/item/escidoc:1587370:1/component/escidoc:1587369/Prb64105.pdf |url-status=dead }}
Use in confocal geometry
Scanning confocal electron energy loss microscopy (SCEELM) is a new analytical microscopy tool that enables a double corrected transmission electron microscope to achieve sub-10 nm depth resolution in depth sectioning imaging of nanomaterials.{{cite journal |last1=Xin |first1=Huolin L. |s2cid=25818886 |display-authors=etal |title=Scanning Confocal Electron Energy-Loss Microscopy Using Valence-Loss Signals|year=2013 |journal=Microscopy and Microanalysis |volume=19|issue=4 |doi=10.1017/S1431927613001438 |pmid=23692691 |bibcode = 2013MiMic..19.1036X |pages=1036–1049}} It was previously termed as energy filtered scanning confocal electron microscopy due to the lack to full spectrum acquisition capability (only a small energy window on the order of 5 eV can be used at a time). SCEELM takes advantages of the newly developed chromatic aberration corrector which allows electrons of more than 100 eV of energy spread to be focused to roughly the same focal plane. It has been demonstrated that a simultaneous acquisition of the zero loss, low-loss, and core loss signals up to 400 eV in the confocal geometry with depth discrimination capability.{{cn|date=February 2024}}
See also
References
{{reflist}}
Further reading
- {{cite book |last=Egerton |first=R. F. |year=1996 |title=Electron Energy Loss Spectroscopy in the Electron Microscope |edition=2nd |location=New York |publisher=Plenum |isbn=978-0-306-45223-9 }}
- {{cite journal |last1=Spence |first1=J. C. H. |author-link1=John C. H. Spence |year=2006 |title=Absorption spectroscopy with sub-angstrom beams: ELS in STEM |journal=Rep. Prog. Phys. |volume=69 |issue=3 |pages=725–758 |doi=10.1088/0034-4885/69/3/R04|bibcode = 2006RPPh...69..725S |s2cid=122148401 }}
- {{cite journal |last1=Gergely |first1=G. |year=2002 |title=Elastic backscattering of electrons: determination of physical parameters of electron transport processes by elastic peak electron spectroscopy |journal=Progress in Surface Science |volume=71 |issue=1 |pages=31–88 |bibcode=2002PrSS...71...31G |doi=10.1016/S0079-6816(02)00019-9}}
- {{cite book |last=Brydson |first=Rik |year=2001 |title=Electron energy loss spectroscopy |publisher=Garland/BIOS Scientific Publishers |isbn=978-1-85996-134-6}}
External links
- [http://www.weels.net A Database of EELS fine structure fingerprints at Cornell]
- [https://eelsdb.eu/ A database of EELS and X-Ray excitation spectra]
- [http://spectrumimager.com/ Cornell Spectrum Imager, an EELS Analysis open-source plugin for ImageJ]
- [http://hyperspy.org HyperSpy, a hyperspectral data analysis Python toolbox especially well suited for EELS data analysis]
- [http://www.eelsmodel.ua.ac.be/ EELSMODEL, software to quantify Electron Energy Loss (EELS) spectra by using model fitting] {{Webarchive|url=https://web.archive.org/web/20170412025325/http://www.eelsmodel.ua.ac.be/ |date=2017-04-12 }}
{{Branches of Spectroscopy}}{{Electron microscopy}}