Modes of mechanical ventilation are one of the most important aspects of the usage of mechanical ventilation. The mode refers to the method of inspiratory support. In general, mode selection is based on clinician familiarity and institutional preferences, since there is a paucity of evidence indicating that the mode affects clinical outcome. The most frequently used forms of volume-limited mechanical ventilation are intermittent mandatory ventilation (IMV) and continuous mandatory ventilation (CMV).{{cite journal | vauthors = Esteban A, Anzueto A, Alía I, Gordo F, Apezteguía C, Pálizas F, Cide D, Goldwaser R, Soto L, Bugedo G, Rodrigo C, Pimentel J, Raimondi G, Tobin MJ | title = How is mechanical ventilation employed in the intensive care unit? An international utilization review | journal = American Journal of Respiratory and Critical Care Medicine | volume = 161 | issue = 5 | pages = 1450–1458 | date = May 2000 | pmid = 10806138 | doi = 10.1164/ajrccm.161.5.9902018 }}
Terminology
There have been substantial changes in the nomenclature of mechanical ventilation over the years, but more recently it has become standardized by many respirology and pulmonology groups.{{cite journal | vauthors = Donn SM | title = Neonatal ventilators: how do they differ? | journal = Journal of Perinatology | volume = 29 | issue = Suppl 2 | pages = S73–S78 | date = May 2009 | pmid = 19399015 | doi = 10.1038/jp.2009.23 | doi-access = free }}{{cite journal | vauthors = Chatburn RL, Volsko TA, Hazy J, Harris LN, Sanders S | title = Determining the basis for a taxonomy of mechanical ventilation | journal = Respiratory Care | volume = 57 | issue = 4 | pages = 514–524 | date = April 2012 | pmid = 22004898 | doi = 10.4187/respcare.01327 | s2cid = 27417478 | url = http://rave.ohiolink.edu/etdc/view?acc_num=ysu1277403555 }}
Writing a mode is most proper in all capital letters with a dash between the control variable and the strategy (i.e. PC-IMV, or VC-MMV etc.).{{citation needed|date=May 2025}}
Taxonomy for mechanical ventilation
The taxonomy is a logical classification system based on 10 maxims of ventilator design:{{cite journal | vauthors = Chatburn RL, El-Khatib M, Mireles-Cabodevila E | title = A taxonomy for mechanical ventilation: 10 fundamental maxims | journal = Respiratory Care | volume = 59 | issue = 11 | pages = 1747–1763 | date = November 2014 | pmid = 25118309 | doi = 10.4187/respcare.03057 | doi-access = free }}
= 10 maxims =
= How modes are classified =
{{How-to|section|date=May 2025}}
Step 1: Identify the primary breath control variable. If inspiration starts with a preset inspiratory pressure, or if pressure is proportional to inspiratory effort, then the control variable is pressure. If inspiration starts with a preset tidal volume and inspiratory flow, then the control variable is volume. If neither is true, the control variable is time.
Step 2: Identify the breath sequence. Determine whether trigger and cycle events are patient or machine determined. Then, use this information to determine the breath sequence.
Step 3: Identify the targeting schemes for the primary breaths and (if applicable) secondary breaths.
= Example mode classification is given below =
Mode Name: A/C Volume Control (Covidien PB 840):{{citation needed|date=March 2020}}
- Inspiratory volume and flow are preset, so the control variable is volume.
- Every breath is volume cycled, which is a form of machine cycling. Any breath for which inspiration is machine cycled is classified as a mandatory breath. Hence, the breath sequence is continuous mandatory ventilation.
- The operator sets all the parameters of the volume and flow waveforms so the targeting scheme is set-point. Thus, the mode is classified as volume control continuous mandatory ventilation with set-point targeting (VC-CMVs).
Mode Name: SIMV Volume Control Plus (Covidien PB 840):{{citation needed|date=March 2020}}
- The operator sets the tidal volume but not the inspiratory flow. Because setting volume alone (like setting flow alone) is a necessary but not sufficient criterion for volume control, the control variable is pressure.
- Spontaneous breaths are allowed between mandatory breaths so the breath sequence is IMV{{clarify|date=March 2020}}.
- The ventilator adjusts inspiratory pressure between breaths to achieve an average preset tidal volume, so the targeting scheme is adaptive. The mode tag is PC-IMVa,s.
Descriptions of common modes
Mechanical ventilation machines are available with both invasive modes (such as intubation) and non-invasive modes (such as BPAP). Invasive has to do with the insertion of medical devices or tubes internal to the patient, while non-invasive is completely external to the patient, as for example in using a tightly fitting mask or other device that covers the patient's nose and mouth.
= Assist mode, control mode, and assist-control mode =
A basic distinction in mechanical ventilation is whether each breath is initiated by the patient (assist mode) or by the machine (control mode).{{Citation needed|date=May 2025}} Dynamic hybrids of the two (assist-control modes) are also possible, and control mode without assist is now mostly obsolete.{{Citation needed|date=May 2025}}
= Airway pressure release ventilation =
File:Airway pressure release ventilation graph.png
Airway pressure release ventilation is a time-cycled alternant between two levels of positive airway pressure, with the main time on the high level and a brief expiratory release to facilitate ventilation.{{Cite journal | vauthors = Stock MC, Downs JB, Frolicher DA | title = Airway Pressure Release Ventilation (APRV): A New Ventilatory Support Mode During Acute Lung Injury (ALI) | journal = Critical Care Medicine | volume = 14 | issue = 4 | pages = 366 | date = April 1986 | doi = 10.1097/00003246-198604000-00111 | url = https://journals.lww.com/ccmjournal/citation/1986/04000/airway_pressure_release_ventilation__aprv___a_new.111.aspx | language = en-US | issn = 1530-0293 | url-access = subscription }}{{Cite web | title = » History of APRV | url = https://www.tcavnetwork.org/history-of-aprv/ | archive-url = http://web.archive.org/web/20250117084044/https://www.tcavnetwork.org/history-of-aprv/ | archive-date = 2025-01-17 | access-date = 2025-03-03 | website = www.tcavnetwork.org | language = en }}{{cite journal | vauthors = Henzler D | title = What on earth is APRV? | journal = Critical Care | volume = 15 | issue = 1 | pages = 115 | year = 2011 | pmid = 21345265 | pmc = 3222047 | doi = 10.1186/cc9419 | publication-place = London, England | doi-access = free }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}}
Airway pressure release ventilation is a very versatile mode of ventilation. Like other modes, it needs to be well understood to use it effectively.{{cite journal | vauthors = Andrews P, Shiber J, Madden M, Nieman GF, Camporota L, Habashi NM | title = Myths and Misconceptions of Airway Pressure Release Ventilation: Getting Past the Noise and on to the Signal | language = English | journal = Frontiers in Physiology | volume = 13 | pages = 928562 | date = 2022-07-25 | pmid = 35957991 | pmc = 9358044 | doi = 10.3389/fphys.2022.928562 | doi-access = free }} The exhalation time (Tlow) is shortened to usually less than one second to maintain alveoli inflation. In the basic sense, this is a continuous pressure with a brief release.
Different perceptions of this mode may exist around the globe. While 'APRV' is common to users in North America, a very similar mode, biphasic positive airway pressure (BIPAP), was introduced in Europe.{{cite journal | vauthors = Baum M, Benzer H, Putensen C, Koller W, Putz G | title = [Biphasic positive airway pressure (BIPAP)--a new form of augmented ventilation] | journal = Der Anaesthesist | volume = 38 | issue = 9 | pages = 452–458 | date = September 1989 | pmid = 2686487 }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}} The term APRV has also been used in American journals where, from the ventilation characteristics, BIPAP would have been perfectly good terminology.{{cite journal | vauthors = Putensen C, Zech S, Wrigge H, Zinserling J, Stüber F, Von Spiegel T, Mutz N | title = Long-term effects of spontaneous breathing during ventilatory support in patients with acute lung injury | journal = American Journal of Respiratory and Critical Care Medicine | volume = 164 | issue = 1 | pages = 43–49 | date = July 2001 | pmid = 11435237 | doi = 10.1164/ajrccm.164.1.2001078 }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}} But BiPAP(tm) is a trademark for a noninvasive ventilation mode in a specific ventilator (Respironics Inc.).
Other manufacturers have followed with their own brand names (BILEVEL, DUOPAP, BIVENT). Although similar in modality, these terms describe how a mode is intended to inflate the lung, rather than defining the characteristics of synchronization or the way spontaneous breathing efforts are supported.
Intermittent mandatory ventilation has not always had the synchronized feature, so the division of modes were understood to be SIMV (synchronized) vs IMV (not-synchronized). Since the American Association for Respiratory Care established a nomenclature of mechanical ventilation the "synchronized" part of the title has been dropped and now there is only IMV.
= Mandatory minute ventilation =
Mandatory minute ventilation (MMV) allows spontaneous breathing with automatic adjustments of mandatory ventilation to the meet the patient's preset minimum minute volume requirement. If the patient maintains the minute volume settings for VT x f, no mandatory breaths are delivered.{{citation needed|date=March 2020}}
If the patient's minute volume is insufficient, mandatory delivery of the preset tidal volume will occur until the minute volume is achieved. The method for monitoring whether or not the patient is meeting the required minute ventilation (VE) differs by ventilator brand and model, but, in general, there is a window of monitored time, and a smaller window checked against the larger window (i.e., in the Dräger Evita® line of mechanical ventilators there is a moving 20-second window, and every 7 seconds the current tidal volume and rate are measured) to decide whether a mechanical breath is needed to maintain the minute ventilation.{{citation needed|date=March 2020}}
MMV is an optimal mode for weaning in neonatal and pediatric populations and has been shown to reduce long-term complications related to mechanical ventilation.{{cite journal | vauthors = Guthrie SO, Lynn C, Lafleur BJ, Donn SM, Walsh WF | title = A crossover analysis of mandatory minute ventilation compared to synchronized intermittent mandatory ventilation in neonates | journal = Journal of Perinatology | volume = 25 | issue = 10 | pages = 643–646 | date = October 2005 | pmid = 16079905 | doi = 10.1038/sj.jp.7211371 | doi-access = free }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}}
= Pressure-regulated volume control =
Pressure-regulated volume control is an Assist Controlled Ventilation (ACV) based mode. Pressure-regulated volume control utilizes pressure-limited, volume-targeted, time-cycled breaths that can be either ventilator- or patient-initiated.{{Citation needed|date=May 2025}}
The peak inspiratory pressure delivered by the ventilator is varied on a breath-to-breath basis to achieve a target tidal volume that is set by the clinician.
For example, if a target tidal volume of 500 mL is set but the ventilator delivers 600 mL, the next breath will be delivered with a lower inspiratory pressure to achieve a lower tidal volume. Though PRVC is regarded as a hybrid mode because of its tidal-volume (VC) settings and pressure-limiting (PC) settings fundamentally PRVC is a pressure-control mode with adaptive targeting.
=Continuous positive airway pressure=
Continuous positive airway pressure (CPAP) is a non-invasive positive pressure mode of respiratory support. CPAP is a continuous pressure applied to keep the alveoli open and not fully deflate. This mechanism for maintaining inflated alveoli helps increase partial pressure of oxygen in arterial blood, an appropriate increase in CPAP increases the PaO2.{{Citation needed|date=May 2025}}
==Automatic positive airway pressure==
Automatic positive airway pressure (APAP) is a form of CPAP that automatically tunes the amount of pressure delivered to the patient to the minimum required to maintain an unobstructed airway on a breath-by-breath basis by measuring the resistance in the patient's breathing.{{Citation needed|date=May 2025}}
= Bilevel positive airway pressure =
Bilevel positive airway pressure (BPAP) is a mode used during non-invasive ventilation (NIV). First used in 1988 by Professor Benzer in Austria,Benzer H (1988) Ventilatory support by intermittent changes in PEEP levels. 4th European Congress on Intensive Care Medicine. Baveno-Stresa it delivers a preset inspiratory positive airway pressure (IPAP) and expiratory positive airway pressure (EPAP). BPAP can be described as a Continuous Positive Airway Pressure system with a time-cycle change of the applied CPAP level.{{cite journal | vauthors = Hörmann C, Baum M, Putensen C, Mutz NJ, Benzer H | title = Biphasic positive airway pressure (BIPAP)--a new mode of ventilatory support | journal = European Journal of Anaesthesiology | volume = 11 | issue = 1 | pages = 37–42 | date = January 1994 | pmid = 8143712 }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}}
CPAP/APAP, BPAP, and other non-invasive ventilation modes have been shown to be effective management tools for chronic obstructive pulmonary disease, acute respiratory failure, sleep apnea, etc.{{cite journal | vauthors = Levitt MA | title = A prospective, randomized trial of BiPAP in severe acute congestive heart failure | journal = The Journal of Emergency Medicine | volume = 21 | issue = 4 | pages = 363–369 | date = November 2001 | pmid = 11728761 | doi = 10.1016/s0736-4679(01)00385-7 }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}}
Often BPAP is incorrectly referred to as "BiPAP". BiPAP is the name of a portable ventilator manufactured by Respironics Corporation; it is just one of many ventilators that can deliver BPAP.{{Citation needed|date=May 2025}}
==Medical uses==
BPAP has been shown to be useful in reducing mortality and reducing the need for endotracheal intubation when used in people with chronic obstructive pulmonary disease (COPD).{{cite journal | vauthors = Osadnik CR, Tee VS, Carson-Chahhoud KV, Picot J, Wedzicha JA, Smith BJ | title = Non-invasive ventilation for the management of acute hypercapnic respiratory failure due to exacerbation of chronic obstructive pulmonary disease | journal = The Cochrane Database of Systematic Reviews | volume = 2017 | issue = 7 | pages = CD004104 | date = July 2017 | pmid = 28702957 | pmc = 6483555 | doi = 10.1002/14651858.CD004104.pub4 | hdl = 10044/1/53458 }}
= High-frequency ventilation (Active) =
The term active refers to the ventilator's forced expiratory system. In a HFV-A scenario, the ventilator uses pressure to apply an inspiratory breath and then applies an opposite pressure to force an expiratory breath. In high-frequency oscillatory ventilation (sometimes abbreviated HFOV) the oscillation bellows and piston force positive pressure in and apply negative pressure to force an expiration.{{cite journal | vauthors = Allardet-Servent J | title = High-frequency oscillatory ventilation in adult patients with acute respiratory distress syndrome: where do we stand and where should we go? | journal = Critical Care Medicine | volume = 39 | issue = 12 | pages = 2761–2762 | date = December 2011 | pmid = 22094505 | doi = 10.1097/CCM.0b013e31822a5c35 }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}}
= High-frequency ventilation (Passive) =
The term passive refers to the ventilator's non-forced expiratory system. In a HFV-P scenario, the ventilator uses pressure to apply an inspiratory breath and then returns to atmospheric pressure to allow for a passive expiration.{{Citation needed|date=May 2025}}
This is seen in High-Frequency Jet Ventilation, sometimes abbreviated HFJV. Also categorized under High Frequency Ventilation is High Frequency Percussive Ventilation, sometimes abbreviated HFPV. With HFPV it utilizes an open circuit to deliver its subtidal volumes by way of the patient interface known as the Phasitron.{{Citation needed|date=May 2025}}
= Volume guarantee =
Volume guarantee an additional parameter available in many types of ventilators that allows the ventilator to change its inspiratory pressure setting to achieve a minimum tidal volume.{{Citation needed|date=May 2025}} This is utilized most often in neonatal patients who need a pressure controlled mode with a consideration for volume control to minimize volutrauma.
Spontaneous breathing and support settings
= Positive end-expiratory pressure =
Positive end expiratory pressure (PEEP) is pressure applied upon expiration. PEEP is applied using either a valve that is connected to the expiratory port and set manually or a valve managed internally by a mechanical ventilator.
PEEP is a pressure that an exhalation has to bypass, in effect causing alveoli to remain open and not fully deflate. This mechanism for maintaining inflated alveoli helps increase partial pressure of oxygen in arterial blood, and an increase in PEEP increases the PaO2.{{cite journal | vauthors = Schuster DP, Klain M, Snyder JV | title = Comparison of high frequency jet ventilation to conventional ventilation during severe acute respiratory failure in humans | journal = Critical Care Medicine | volume = 10 | issue = 10 | pages = 625–630 | date = October 1982 | pmid = 6749433 | doi = 10.1097/00003246-198210000-00001 }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}}
= Pressure support =
Pressure support is a spontaneous mode of ventilation also named Pressure Support Ventilation (PSV). The patient initiates every breath and the ventilator delivers support with the preset pressure value. With support from the ventilator, the patient also regulates their own respiratory rate and their tidal volume.
In Pressure Support, the set inspiratory pressure support level is kept constant and there is a decelerating flow. The patient triggers all breaths. If there is a change in the mechanical properties of the lung/thorax and patient effort, the delivered tidal volume will be affected. The user must then regulate the pressure support level to obtain desired ventilation.{{cite web | title = Servo-i Mechanical Ventilator | url = https://www.getinge.com/int/products/servo-i-mechanical-ventilator/ | access-date = 17 May 2025 | work = Getinge AB }}{{cite web | title = Servo-s Mechanical Ventilator | url = https://www.getinge.com/int/products/servo-s-mechanical-ventilator/ | access-date = 17 May 2025 | work = Getinge AB }}
Pressure support improves oxygenation,{{cite journal | vauthors = Spieth PM, Carvalho AR, Güldner A, Kasper M, Schubert R, Carvalho NC, Beda A, Dassow C, Uhlig S, Koch T, Pelosi P, Gama de Abreu M | title = Pressure support improves oxygenation and lung protection compared to pressure-controlled ventilation and is further improved by random variation of pressure support | journal = Critical Care Medicine | volume = 39 | issue = 4 | pages = 746–755 | date = April 2011 | pmid = 21263322 | doi = 10.1097/CCM.0b013e318206bda6 | s2cid = 35876431 }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}} ventilation and decreases work of breathing.
Also see adaptive support ventilation.
Other ventilation modes and strategies
=Flow-controlled ventilation=
Flow-controlled ventilation (FCV) is an entirely dynamic ventilation mode, without pauses, with continuous and stable gas flows during both inspiration and expiration, aiming for linear changes in both volume and pressure.{{cite patent | country = DE | number = 102016109528A1 | inventor = Enk D | title = Verfahren und Vorrichtung zur Beatmung eines Patienten (method and device for ventilating a patient) | assign = Current Assignee Ventinova Technologies BV | pubdate = 7 September 2017 | url = https://patents.google.com/patent/DE102016109528A1/de }} FCV is an invasive ventilation mode but, unlike Volume- and pressure controlled modes, it does not rely on a passive expiration created by collapse of the thoracic wall and elastic recoil of the lungs. A high resistant breathing circuit inhibits a passive expiration and therewith allows to fully control and stabilize the expiration flow. FCV creates an inspiration by generating a stable flow from a set End-expiratory pressure (EEP) to a set Peak pressure. Then a stable expiratory flow is created by suctioning.{{cite patent | inventor = Enk D | title = Gasstromumkehrelement (gas flow reversing element). | country = DE | number = 102007013385A1 | pubdate = 18 September }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}} This expiratory flow rate is preferably similar to the inspiratory flow, aiming for an I:E ratio of 1:1.0, to minimize energy dissipation in the lungs.{{cite journal | vauthors = Barnes T, van Asseldonk D, Enk D | title = Minimisation of dissipated energy in the airways during mechanical ventilation by using constant inspiratory and expiratory flows - Flow-controlled ventilation (FCV) | journal = Medical Hypotheses | volume = 121 | issue = | pages = 167–176 | date = December 2018 | pmid = 30396474 | doi = 10.1016/j.mehy.2018.09.038 | doi-access = free }}{{Medical citation needed|date=May 2025}}{{cite journal | vauthors = Barnes T, Enk D |title=Ventilation for low dissipated energy achieved using flow control during both inspiration and expiration |journal=Trends in Anaesthesia and Critical Care |date=February 2019 |volume=24 |pages=5–12 |doi=10.1016/j.tacc.2018.09.003|doi-access=free }} FCV is a more efficient ventilation as compared to conventional modes,{{cite journal | vauthors = Weber J, Schmidt J, Straka L, Wirth S, Schumann S | title = Flow-controlled ventilation improves gas exchange in lung-healthy patients- a randomized interventional cross-over study | journal = Acta Anaesthesiologica Scandinavica | volume = 64 | issue = 4 | pages = 481–488 | date = April 2020 | pmid = 31828755 | doi = 10.1111/aas.13526 | url = | doi-access = free }}{{cite journal | vauthors = Sebrechts T, Morrison SG, Schepens T, Saldien V | title = Flow-controlled ventilation with the Evone ventilator and Tritube versus volume-controlled ventilation: A clinical cross-over pilot study describing oxygenation, ventilation and haemodynamic variables | journal = European Journal of Anaesthesiology | volume = 38 | issue = 2 | pages = 209–211 | date = February 2021 | pmid = 33394805 | doi = 10.1097/EJA.0000000000001326 | url = }}{{Primary source inline|date=May 2025}} allows ventilation through even small lumens (~2 – 10 mm ID){{cite journal | vauthors = Schmidt J, Günther F, Weber J, Kehm V, Pfeiffer J, Becker C, Wenzel C, Borgmann S, Wirth S, Schumann S | title = Glottic visibility for laryngeal surgery: Tritube vs. microlaryngeal tube: A randomised controlled trial | journal = European Journal of Anaesthesiology | volume = 36 | issue = 12 | pages = 963–971 | date = December 2019 | pmid = 31644514 | pmc = 6855316 | doi = 10.1097/EJA.0000000000001110 | url = }}{{cite journal | vauthors = Meulemans J, Jans A, Vermeulen K, Vandommele J, Delaere P, Vander Poorten V | title = Evone® Flow-Controlled Ventilation During Upper Airway Surgery: A Clinical Feasibility Study and Safety Assessment | journal = Frontiers in Surgery | volume = 7 | issue = | pages = 6 | date = 2020 | pmid = 32185179 | pmc = 7058692 | doi = 10.3389/fsurg.2020.00006 | doi-access = free | url = }}{{Primary source inline|date=May 2025}} and results in less applied mechanical power. FCV was invented by Professor Dr. med. Dietmar Enk.{{Primary source inline|date=May 2025|reason=WP:PATENTS are generally poor sources}}
=Negative pressure ventilation=
:Main article: Negative pressure ventilator
Negative-pressure ventilation stimulates (or forces) breathing by periodic application of partial vacuum (air pressure reduced below ambient pressure), applied externally to the patient's torso—specifically, chest and abdomen—to assist (or force) the chest to expand, expanding the lungs, resulting in voluntary (or involuntary) inhalation through the patient's airway.{{cite journal | vauthors = Matioc AA | title = An Anesthesiologist's Perspective on the History of Basic Airway Management: The "Progressive" Era, 1904 to 1960 | journal = Anesthesiology | volume = 128 | issue = 2 | pages = 254–271 | date = February 2018 | pmid = 29112511 | doi = 10.1097/ALN.0000000000001975 | quote = Early Positive and Alternate Pressure Machines }}Rockoff, Mark, M.D., [https://www.youtube.com/watch?v=cUCqsl6JROg "The Iron Lung and Polio,"], video (8 minutes), January 11, 2016, OPENPediatrics and Boston Children's Hospital on YouTube, retrieved April 11, 2020 (historical background and images, explanatory diagrams, and live demonstrations){{cite book | vauthors = Walkey A, Summer R | quote = Negative pressure | chapter = E. Noninvasive Mechanical Ventilation | url = https://www.bumc.bu.edu/im-residency/files/2010/10/Residents-Critical-Care-Handbook.pdf | title = Boston Medical Center ICU Manual | date = 2008 | publisher = Boston University | pages = 17 }}
Various "negative pressure ventilators" (NPVs) have been developed to serve this function—most famously the "Iron lung," a tank in which the patient lays, with only their head exposed to ambient air, while air pressure on the remainder of their body, inside the tank, is varied by pumping, to stimulate chest and lung expansion and contraction. Though not in wide use today, NPVs were the principal forms of hospital and long-term mechanical ventilation in the first half of the 20th century, and remain in limited use today.
= Closed loop systems =
{{cs1 config|name-list-style=vanc|display-authors=6}}
==Adaptive Support Ventilation==
Adaptive Support Ventilation (ASV) is the only commercially available mode that uses optimal targeting.{{cite book | vauthors = Tehrani FT | chapter = Automatic Control of an Artificial Respirator | title = Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society | volume = 13 | pages = 1738–1739 | date = October 1991 | doi = 10.1109/IEMBS.1991.684729 | isbn = 978-0-7803-0216-7 | s2cid = 63221714 }}{{cite journal | vauthors = Chatburn RL, Mireles-Cabodevila E | title = Closed-loop control of mechanical ventilation: description and classification of targeting schemes | journal = Respiratory Care | volume = 56 | issue = 1 | pages = 85–102 | date = January 2011 | pmid = 21235841 | doi = 10.4187/respcare.00967 }} In this positive pressure mode of ventilation, the frequency and tidal volume of breaths of a patient on the ventilator are automatically adjusted and optimized to mimic natural breathing, stimulate spontaneous breathing, and reduce weaning time. In the ASV mode, every breath is synchronized with patient effort if such an effort exists, and otherwise, full mechanical ventilation is provided to the patient.{{cite journal | vauthors = Tehrani FT | title = Automatic control of mechanical ventilation. Part 1: theory and history of the technology | journal = Journal of Clinical Monitoring and Computing | volume = 22 | issue = 6 | pages = 409–415 | date = December 2008 | pmid = 19011976 | doi = 10.1007/s10877-008-9150-z }}{{cite journal | vauthors = Tehrani FT | title = Automatic control of mechanical ventilation. Part 2: the existing techniques and future trends | journal = Journal of Clinical Monitoring and Computing | volume = 22 | issue = 6 | pages = 417–424 | date = December 2008 | pmid = 19020981 | doi = 10.1007/s10877-008-9151-y }}
==Automatic Tube Compensation==
Automatic Tube Compensation (ATC) is the simplest example of a computer-controlled targeting system on a ventilator. It is a form of servo targeting.{{Citation needed|date=May 2025}}
The goal of ATC is to support the resistive work of breathing through the artificial airway.
==Neurally Adjusted Ventilatory Assist==
Neurally Adjusted Ventilatory Assist (NAVA) is adjusted by a computer (servo) and is similar to ATC but with more complex requirements for implementation.{{Citation needed|date=May 2025}}
In terms of patient-ventilator synchrony, NAVA supports both resistive and elastic work of breathing in proportion to the patient's inspiratory effort
==Proportional Assist Ventilation==
Proportional assist ventilation (PAV) is another servo targeting based mode in which the ventilator guarantees the percentage of work regardless of changes in pulmonary compliance and resistance.{{cite journal | vauthors = Younes M | title = Proportional assist ventilation, a new approach to ventilatory support. Theory | journal = The American Review of Respiratory Disease | volume = 145 | issue = 1 | pages = 114–120 | date = January 1992 | pmid = 1731573 | doi = 10.1164/ajrccm/145.1.114 }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}}
The ventilator varies the tidal volume and pressure based on the patient's work of breathing. The amount it delivers is proportional to the percentage of assistance it is set to give.
PAV, like NAVA, supports both restrictive and elastic work of breathing in proportion to the patient's inspiratory effort.
= Liquid ventilation =
Liquid ventilation is a technique of mechanical ventilation in which the lungs are insufflated with an oxygenated perfluorochemical liquid rather than an oxygen-containing gas mixture. The use of perfluorochemicals, rather than nitrogen, as the inert carrier of oxygen and carbon dioxide offers a number of theoretical advantages for the treatment of acute lung injury, including:
- Reducing surface tension by maintaining a fluid interface with alveoli
- Opening of collapsed alveoli by hydraulic pressure with a lower risk of barotrauma
- Providing a reservoir in which oxygen and carbon dioxide can be exchanged with pulmonary capillary blood
- Functioning as a high-efficiency heat exchanger
Despite its theoretical advantages, efficacy studies have been disappointing and the optimal clinical use of LV has yet to be defined.{{cite journal | vauthors = Degraeuwe PL, Vos GD, Blanco CE | title = Perfluorochemical liquid ventilation: from the animal laboratory to the intensive care unit | journal = The International Journal of Artificial Organs | volume = 18 | issue = 10 | pages = 674–683 | date = October 1995 | pmid = 8647601 | doi = 10.1177/039139889501801020 | s2cid = 13038566 }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}}
== Total liquid ventilation ==
In total liquid ventilation (TLV), the entire lung is filled with an oxygenated PFC liquid, and a liquid tidal volume of PFC is actively pumped into and out of the lungs. A specialized apparatus is required to deliver and remove the relatively dense, viscous PFC tidal volumes, and to extracorporeally oxygenate and remove carbon dioxide from the liquid.{{cite journal | vauthors = Norris MK, Fuhrman BP, Leach CL | title = Liquid ventilation: it's not science fiction anymore | journal = AACN Clinical Issues in Critical Care Nursing | volume = 5 | issue = 3 | pages = 246–254 | date = August 1994 | pmid = 7780839 | doi = 10.4037/15597768-1994-3004 }}{{cite journal | vauthors = Greenspan JS | title = Physiology and clinical role of liquid ventilation therapy | journal = Journal of Perinatology | volume = 16 | issue = 2 Pt 2 Su | pages = S47–S52 | year = 1996 | pmid = 8732549 }}{{cite journal | vauthors = Dirkes S | title = Liquid ventilation: new frontiers in the treatment of ARDS | journal = Critical Care Nurse | volume = 16 | issue = 3 | pages = 53–58 | date = June 1996 | pmid = 8852261 | doi = 10.4037/ccn1996.16.3.53 }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}}
== Partial liquid ventilation ==
In partial liquid ventilation (PLV), the lungs are slowly filled with a volume of PFC equivalent or close to the FRC during gas ventilation. The PFC within the lungs is oxygenated and carbon dioxide is removed by means of gas breaths cycling in the lungs by a conventional gas ventilator.{{cite journal | vauthors = Cox CA, Wolfson MR, Shaffer TH | title = Liquid ventilation: a comprehensive overview | journal = Neonatal Network | volume = 15 | issue = 3 | pages = 31–43 | date = April 1996 | pmid = 8715647 }}{{Update inline|date=May 2025|?=yes|reason=Source fails WP:MEDDATE. Is this still true?}}
See also
{{Portal|Medicine}}
- {{annotated link|Table of modes of mechanical ventilation}}
- {{annotated link|Mechanical ventilation}}
- Prone ventilation
- {{annotated link|Respiratory therapist}}
- {{annotated link|Bubble CPAP}}
References
{{reflist|30em}}
{{Mechanical ventilation}}
Category:Intensive care medicine